Key result
Point-of-care skin ultrasound demonstrates excellent reliability for measuring subcutaneous tissue thickness in neonates and infants.
Why the study?
Fluid overload is common after pediatric cardiac surgery and linked to morbidity and mortality, but no gold standard exists for assessing fluid status.
Does point-of-care ultrasound reliably measure subcutaneous tissue thickness to quantify extravascular fluid overload in neonates and infants undergoing cardiac surgery?
Cohort (n=72)
No
Does point-of-care ultrasound reliably measure subcutaneous tissue thickness to quantify extravascular fluid overload in neonates and infants undergoing cardiac surgery?
Point-of-care skin ultrasound is a reproducible and reliable non-invasive method to quantify extravascular fluid overload (edema) in neonates and infants following cardiac surgery.
Answers are grounded in this paper and cite their sources.
POCUS skin edema measurement may aid postoperative fluid assessment in infants; leaves open clinical adoption pending outcome validation.
Purpose: Fluid overload is a common post-operative issue in children following cardiac surgery and is associated with increased morbidity and mortality. There is currently no gold standard for evaluating fluid status. We sought to validate the use of point-of-care ultrasound to measure skin edema in infants and assess the intra- and inter-user variability. Methods: Prospective cohort study of neonates (≤30 d/o) and infants (31 d/o to 12 m/o) undergoing cardiac surgery and neonatal controls. Skin ultrasound was performed on four body sites at baseline and daily post-operatively through post-operative day (POD) 3. Subcutaneous tissue depth was manually measured. Intra- and inter-user variability was assessed using intraclass correlation coefficient (ICC). Results: Fifty control and 22 surgical subjects underwent skin ultrasound. There was no difference between baseline surgical and control neonates. Subcutaneous tissue increased in neonates starting POD 1 with minimal improvement by POD 3. In infants, this pattern was less pronounced with near resolution by POD 3. Intra-user variability was excellent (ICC 0.95). Inter-user variability was very good (ICC 0.82). Conclusion: Point-of-care skin ultrasound is a reproducible and reliable method to measure subcutaneous tissue in infants with and without congenital heart disease. Acute increases in subcutaneous tissue suggests development of skin edema, consistent with extravascular fluid overload. There is evidence of skin edema starting POD 1 in all subjects with no substantial improvement by POD 3 in neonates. Point-of-care ultrasound could be an objective way to measure extravascular fluid overload in infants. Further research is needed to determine how extravascular fluid overload correlates to clinical outcomes.
No takes yet. Share an insight, caveat, or question.
Persson et al. (2021) conducted a cohort in Congenital heart disease (n=72). Point-of-care skin ultrasound vs. Healthy controls / Baseline measurements was evaluated on Intra-user variability of subcutaneous tissue thickness measurements (ICC 0.95, 95% CI 0.946-0.954). Point-of-care skin ultrasound demonstrated excellent intra-user reliability and very good inter-user reliability for measuring subcutaneous tissue thickness in neonates and infants.