An instrument for measurement of the profiles of optically rough surfaces is described. The sensor is a robust and compact Fizeau interferometer, in which one reflection is derived from the test surface. A laser-diode source and a photodetector communicate with the sensor through an optical fiber. The optical output is demodulated with a phase-stepping algorithm, achieved by frequency modulating the source. The measured horizontal resolution was 7 μm, and the noise-limited vertical resolution was 0.3 nm/Hz.
No takes yet. Share an insight, caveat, or question.
Hand et al. (1993) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: