Key result
Rapid TST-blot yields 100% concordance with standard 3-day immunodiffusion for detecting toxic shock syndrome toxin.
Why the study?
Does the TST-blot assay accurately and rapidly detect TST production in S. aureus isolates compared to the Ouchterlony immunodiffusion assay?
Does the TST-blot assay accurately and rapidly detect TST production in S. aureus isolates compared to the Ouchterlony immunodiffusion assay?
Effect estimate: 100% concordance
Absolute Event Rate: 37.6% vs 37.6%
The TST-blot is a rapid and accurate method for detecting toxic shock syndrome toxin in S. aureus isolates, suitable for diagnostic and epidemiological studies.
May accelerate TST screening in S. aureus isolates; leaves open prospective clinical validation before diagnostic adoption.
A rapid immunoblot assay (TST-blot) was developed and used to screen Staphylococcus aureus isolates for toxic shock syndrome toxin (TST) production. Growth from an 18-h stab inoculum of S. aureus on brain heart infusion agar was transferred directly to a nitrocellulose sheet. Nonspecific protein binding sites were blocked with bovine serum albumin, and the nitrocellulose sheet was incubated with affinity-purified antibody to TST, followed by incubation with horseradish peroxidase-conjugated protein A. Toxin was visualized by detection of the peroxidase-conjugated protein A-anti TST-TST complex with 4-chloro-1-napthol. The sensitivities and specificities of the TST-blot and Ouchterlony microslide immunodiffusion assay were compared by screening 141 S. aureus isolates for TST production. In both assays, 53 of 141 isolates produced detectable levels of TST, whereas 88 isolates produced no toxin. A 100% concordance was found between the two assays. The TST-blot yielded the same results in less than 24 h as those yielded by the 3-day immunodiffusion assay. Thus, this rapid method for detection of TST in multiple samples appears to be well suited for diagnostic and epidemiological studies. Furthermore, it would appear to be ideal for use in TST genetics research.
No takes yet. Share an insight, caveat, or question.
Weckbach et al. (1984) studied Staphylococcus aureus isolates (toxic shock syndrome toxin production) (n=141). TST-blot (rapid immunoblot assay) vs. Ouchterlony microslide immunodiffusion assay was evaluated on Detection of toxic shock syndrome toxin (TST) production (100% concordance). The rapid TST-blot assay demonstrated 100% concordance with the standard 3-day immunodiffusion assay in detecting toxic shock syndrome toxin production among 141 S. aureus isolates.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: