Demonstrates that optimal CT reconstruction phases depend on heart rate in patients undergoing thoracic aorta imaging, indicating tailored timing minimizes motion artifact.
The authors assessed motion artifact of the thoracic aorta in 25 patients who underwent multi-detector row computed tomography (CT) with retrospective electrocardiographic (ECG) gating. CT reconstructions centered at four phases of diastole were compared for five different levels of the thoracic aorta. A significant positive correlation was observed between heart rate and motion artifact (r = 0.72, P <.001). The optimal reconstruction phase varied between patients, and this was directly related to heart rate. For patients with a heart rate of 70 beats per minute, the reconstruction phase centered at 75% of the R-R interval had the significantly least motion artifact (P =.004). Conversely, the optimal reconstruction phase for patients with heart rates above 70 beats per minute was centered at 50% of the R-R interval (P =.09).
No takes yet. Share an insight, caveat, or question.
Morgan‐Hughes et al. (2003) studied this question.