Key result
Point-of-care PATHFAST hs-cTnI correlates strongly with standard Centaur assays and reduces total turnaround time.
Why the study?
Cardiac troponin I is a useful biomarker for acute myocardial infarction, but there is a wide variety of point-of-care analysers providing its measurement.
Does a point-of-care high-sensitivity cTnI assay on PATHFAST provide comparable results to central laboratory assays with improved turnaround time?
Observational (n=161)
Does a point-of-care high-sensitivity cTnI assay on PATHFAST provide comparable results to central laboratory assays with improved turnaround time?
The PATHFAST point-of-care high-sensitivity troponin assay provides comparable results to central laboratory assays while reducing turnaround time in the emergency room.
Supports the clinical utility of point-of-care high-sensitivity troponin testing for rapid triage.
Aim: Although the levels of cardiac troponin I (cTnI) have proved to be a useful diagnostic biomarker of acute myocardial infarction, there are a wide variety of point-of-care (POC) analysers, which provide measurements of cTnI. The aim of this study was to compare the results obtained by the ADVIA Centaur ultra-assay cTnI assay (us-cTnI), ADVIA Centaur high-sensitive cTnI assay (hs-cTnI) and a POC high-sensitivity assay using PATHFAST. We also aimed to explore total turnaround time (TAT) for laboratory results using the POC PATHFAST analyser. Methods: Samples from 161 patients were taken. Of these samples, 129 were tested with all three assays (us-cTnI, hs-cTnI and PATHFAST), and 32 samples were tested on PATHFAST for the comparison of whole blood, serum and plasma. Results: Comparison of the POC testing methods in this study demonstrated that there are strong linear relationships between all three cTnI assays (us-cTnI, hs-cTnI and POC on PATHFAST). Furthermore, we also show there are strong linear relationships between the two high-sensitive cTnI assays (hs-cTnI and PATHFAST) for blood serum samples, as determined by Passing-Bablok regression analyses. In our comparison of our new data with our older study, the TAT went down. Conclusion: The timeliness of laboratory results is, in addition to accuracy and precision, one of the key indicators of laboratory performance, and at the same time has a significant impact on the course of the patient's condition. It is therefore important that the laboratory strives to meet the expectations of clinicians regarding the time from the order to the result of the analysis.
No takes yet. Share an insight, caveat, or question.
Osredkar et al. (2021) conducted an observational in Acute myocardial infarction (n=161). POC high-sensitivity assay using PATHFAST vs. ADVIA Centaur us-cTnI and hs-cTnI assays was evaluated on Linear relationships between cTnI assays and total turnaround time. A point-of-care high-sensitivity assay using PATHFAST demonstrated strong linear relationships with ADVIA Centaur us-cTnI and hs-cTnI assays, and reduced total turnaround time.