Key result
Pathfast cTnI assay establishes a 99th percentile upper reference limit of 0.013 mg/L in healthy individuals.
Why the study?
Cardiac troponins are essential for risk stratification and diagnosis in ACS, but analytical performance data for the Pathfast cTnI method were needed following CLSI guidelines.
Observational (n=320)
Yes
The Pathfast cTnI assay demonstrates analytical performance characteristics suitable for clinical use in the risk stratification and diagnosis of acute coronary syndrome.
Provides reference values for Pathfast cTnI interpretation; leaves open prospective validation of diagnostic performance in ACS.
BACKGROUND: Cardiac troponins are considered the cornerstone for risk stratification and diagnosis of patients with acute coronary syndrome (ACS). Following Clinical Laboratory Standards Institute (CLSI) guidelines, we assessed the analytical performances of the Pathfast (Mitsubishi, Japan) cTnI method. METHODS: We evaluated different sample types. Control materials and lithium heparin plasma pools were used to determine: limit of blank (LoB), limit of detection (LoD), imprecision and linearity. The effects of potential endogenous interfering substances and the possibility of falsely increased cardiac troponin I (cTnI) concentrations attributable to the presence of heterophilic antibodies (HA), rheumatoid factor (RF) and human anti-mouse antibodies (HAMA) in high concentrations were evaluated. The 99th percentile limit of the cTnI value distribution was determined from 320 Caucasian reference individuals. RESULTS: No significant differences were found when cTnI concentrations of 40 lithium-heparin plasma samples were compared with the matched values of K(2)-EDTA plasma, whole blood and serum samples. The LoB and the LoD of the cTnI method were 0.0048 and 0.0066 microg/L, respectively. cTnI mean values from 0.66 to 6.0 microg/L showed a total CV% from 6.0 to 6.4. cTnI at a concentration of 0.02 microg/L was associated with a total CV of 9.6%. The method gave a linear response for cTnI concentrations within the measurement range. In six of 12 samples containing HA, a positive interference was demonstrated. The 99th percentile limit of the cTnI distribution in the reference population was 0.013 microg/L. CONCLUSIONS: The data indicate that the cTnI Pathfast method may be suitable for helping clinicians in the management of patients with ACS.
No takes yet. Share an insight, caveat, or question.
Serio et al. (2009) conducted an observational in Healthy subjects (for reference limit) and Acute Coronary Syndrome (n=320). Pathfast cardiac troponin I assay vs. Stratus CS cTnI method was evaluated on 99th percentile limit of the cTnI distribution in the reference population (95% CI 0.0068-0.019). The Pathfast cardiac troponin I assay demonstrated a 99th percentile upper reference limit of 0.013 mg/L in a healthy population, with a limit of detection of 0.0066 mg/L.