Answers are grounded in this paper and cite their sources.
A new sampling approach has been developed to enable affordable, time-resolved monitoring of particulate chemical compositions, and more generally to provide concentrated samples of airborne particles. Using a newly developed, moderated water-based condensational growth technology, individual particle samples are deposited in a 1-mm diameter dry "spot". The moderated condensation technology enables this collection with minimal temperature rise, providing robust collection for volatile constituents. Measured collection efficiencies are above 95% for particles in the size range from 0.010 μm to 2.5 μm. A set of 20 or more time-resolved samples, plus blanks, may be collected onto a multiwell collection plate. For chemical analysis the plate is returned to the laboratory, and placed directly into a modified autosampler, without extraction or preparation. The autosampler handles the addition of eluent, extraction, and sample injection without user manipulation. This paper presents the design and laboratory evaluation of a 1.5 L/min sampling rate version of this system.
No takes yet. Share an insight, caveat, or question.
Fernandez et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: