Experimental study demonstrates magnetic-field-tunable resonant detection in indium antimonide above 1 THz, indicating limits for tunable spectroscopy.
Responsivity of a bulk InSb detector at liquid helium temperature was studied in the frequency range 0.1 THz < f < 2.5 THz.The detector was exposed to a monochromatic radiation and its response was registered as a function of magnetic field up to 7 T. Due to a very strong magnetoresistance of InSb, a dependence of the photocurrent on magnetic field shows a totally different shape than that of the photoresistance.Thus, photoconductivity was found to be the only reliable way of data presentation.The detector develops a resonant response, related to the cyclotron resonance absorption, at f > ∼ 0.5 THz but this response dominates the spectrum only at f > ∼ 1 THz.At f < ∼ 1 THz the response is mainly non-resonant and is caused by intra-band transitions.We conclude that an InSb detector can be used as a magnetic-field-tunable spectroscopic element only if one can assure that the frequency of photons which reach the detector is higher than about 1 THz.