Key result
Three commercial HDV RNA assays show 100% qualitative concordance despite variability in viral load quantification.
Why the study?
Precise HDV-RNA detection and quantification are pivotal for accurate diagnosis and monitoring response to newly approved treatment.
Cross-Sectional (n=151)
Three commercial HDV RNA assays demonstrate good qualitative concordance and quantitative correlation, but lack of harmonization requires monitoring to be performed using the same assay and laboratory.
Variability in HDV RNA quantification warrants caution in clinical monitoring; leaves open the need for assay harmonization.
Precise HDV-RNA detection and quantification are pivotal for accurate diagnosis and monitoring of response to newly approved treatment. We evaluate the performance of three assays: Hepatitis Delta RT-PCR system, EurobioPlex HDV assay, and the RoboGene HDV RNA Quantification kit 2.0, testing 151 HBsAg-positive samples, 90 HDV-RNA negative and 61 HDV-RNA positive. We also evaluated serial dilutions of the WHO international standard for HDV, PEI 7657/12. Finally, all HDV-RNA positive samples were genotyped using a next-generation sequencing strategy. Qualitative results indicated a 100% concordance between tests. Quantitative results correlated well between tests, r2=0.703 (Vircell-vs-Eurobio), r2=0.833 (Vircell-vs-RoboGene), r2=0.835 (Robogene-vs-Eurobio). Bias index was 2.083 (Vircell-vs-Eurobio), -1.283 (Vircell-vs-RoboGene), and -3.36 (Robogene-vs-Eurobio). Using the WHO IS, Vircell overestimated the viral load by 0.98 log IU/mL, Eurobio by 1.46 log IU/mL, and RoboGene underestimated it by 0.98 log IU/mL. Fifty-nine samples were successfully genotyped (Genotype 1, n=52; Genotype 5, n=7; Genotype 6, n=1), with similar results in terms of correlation and bias. The three assays demonstrate promising performance. Their accuracy, reflected in concordance rates and quantitative analysis, makes them suitable for clinical practice. However, in the absence of assay harmonization, quantitative HDV RNA monitoring should be performed in the same laboratory and assay to avoid variability.
No takes yet. Share an insight, caveat, or question.
Illescas-López et al. (2024) conducted a cross-sectional in Hepatitis D (n=151). Three commercial HDV RNA detection assays vs. Each other and WHO international standard was evaluated on Qualitative concordance and quantitative correlation. Three commercial HDV RNA assays demonstrated 100% qualitative concordance and good quantitative correlation (r2=0.703 to 0.835), though lack of harmonization causes viral load variability.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: