Answers are grounded in this paper and cite their sources.
In clinical MR neuroimaging, 3D fluid-attenuated inversion recovery (3D-FLAIR) with a variable-flip-angle turbo spin echo sequence is becoming popular. There are more than 100 reports regarding 3D-FLAIR in the PubMed database. In this article, the technical and clinical features of 3D-FLAIR for neuroimaging are reviewed and summarized. 3D-FLAIR allows thinner slices with multi-planar reformation capability, a higher flow sensitivity, high sensitivity to subtle T1 changes in fluid, images without cerebrospinal fluid (CSF) inflow artifacts, and a 3D dataset compatible with computer-aided analysis. In addition, 3D-FLAIR can be obtained within a clinically reasonable scan time. It is important for radiologists to be familiar with the features of 3D-FLAIR and to provide useful information for patients.
No takes yet. Share an insight, caveat, or question.
Shinji Naganawa (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: