This is the authors' abstract. We don't add key points for this paper.
Reaction history measurements are fundamental components of diagnosing inertial confinement fusion (ICF) implosions, which require implementation and development on measuring fusion gamma rays. A new detection system of fusion gamma ray, which performs an ultrafast time response, has been developed recently. Using a pulsed gamma-ray source with a duration of about 10 ps, a temporal resolution of better than 100 ps was determined by two methods in the experiment. In addition, the detection system shows a reliable merit in radiation shielding, which is resulted from the advantages of the laser probe. A flat feature of the energy response, over the range of higher than ∼2 MeV, has also been acquired with Monte Carlo methods. This detection system exhibits a temporal response of the rising time of ∼70 ps and the full width at half maximum of ∼117 ps, which is very promising for some ICF reaction history measurements.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Tan et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: