Answers are grounded in this paper and cite their sources.
We have developed a dual goniometer end-station using Stanford Synchrotron Radiation Lightsource, beamline 16-2, in the energy range of 650–2400 eV for crystal calibrations. Three alkali acid phthalate crystals [potassium acid phthalate (KAP), rubidium acid phthalate (RbAP), and cesium acid phthalate (CsAP)] with (001) Miller indices are used to demonstrate the capability of our current experimental setup. The beamline provides photon fluxes from about 10 9 to 10 11 photons/ sec between 50 and 2400 eV, with an error less than 1% absolute energy, and energy resolution between 1250 and 1800 depending on the energy. Our crystal calibrations are limited to energies above 650 eV due to our selection of an Amptek 8 µm Be window fast SSD detector. Total reflectivity calibrations of the crystals can be performed by either angle or photon energy scans. Repeatability of rocking curve measurements is better than 2%. Calibration results show significant impacts of Rb L and Cs M absorption edges for reflectivity in this energy range. The results are compared with XOP calculations, a publicly available multi-lamellar (ML) model, with good agreement. This indicates that it is possible to predict performances of flat KAP and RbAP crystals in this energy range when using the ML model.
No takes yet. Share an insight, caveat, or question.
Conlon et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: