Key result
Harmonic imaging outperforms conventional echo for micro vascular plug visualization, enabling fluoroscopy-free deployment.
Why the study?
The Medtronic micro vascular plug is poorly visible on fluoroscopy and echocardiography during off-label transcatheter occlusion of patent ductus arteriosus in premature infants.
Does harmonic imaging improve the visualization and deployment of the Medtronic micro vascular plug compared to conventional echocardiography in premature infants with patent ductus arteriosus?
Observational (n=9)
Does harmonic imaging improve the visualization and deployment of the Medtronic micro vascular plug compared to conventional echocardiography in premature infants with patent ductus arteriosus?
Harmonic echocardiographic imaging allows for precise, fluoroscopy-free implantation of the Medtronic micro vascular plug for PDA closure in premature infants.
Answers are grounded in this paper and cite their sources.
May support fluoroscopy-free PDA closure in premature infants; hypothesis-generating and requires prospective validation before practice change.
OBJECTIVES: To report the usefulness of harmonic imaging in echocardiography to visualize and direct the implantation of the Medtronic micro vascular plug (MVP). BACKGROUND: Off label use of the MVP was reported for transcatheter occlusion of patent ductus arteriosus (PDA) in premature infants. The device is poorly visible on fluoroscopy and echocardiography. METHODS: In 9 consecutive premature infants, the MVP was used for transcatheter closure of the PDA. In each, the ability of conventional echocardiographic imaging was compared to harmonic imaging, and the device was deployed in the PDA using echocardiography. RESULTS: In each subject, harmonic imaging proved superior to conventional echocardiography to visualize the MVP in premature infants using 12 and 8 MHz probes. Once the delivery, catheter was across the PDA into the descending aorta, and the MVP advanced to the catheter tip, positioning, and deployment of the device was possible without fluoroscopy. All devices were deployed appropriately with immediate occlusion and no obstruction to the left pulmonary artery or aorta. CONCLUSIONS: The MVP can be accurately imaged using harmonic imaging, even in the near field in premature infants. Precise implantation of the MVP in the PDA of premature infants is possible with echocardiographic imaging of the device and vascular structures.
No takes yet. Share an insight, caveat, or question.
Berry et al. (2019) conducted an observational in Patent ductus arteriosus (n=9). Harmonic imaging in echocardiography vs. Conventional echocardiographic imaging was evaluated on Visualization and deployment of the Medtronic micro vascular plug. Harmonic imaging was superior to conventional echocardiography for visualizing the Medtronic micro vascular plug, enabling successful deployment without fluoroscopy in 9 premature infants.