Key result
Fourth-generation FloTrac/Vigileo overestimates cardiac output by ~0.94 L/min vs continuous thermodilution during OPCAB.
Why the study?
Intermittent thermodilution has errors and measurement limitations when used as a reference for FloTrac/Vigileo cardiac output monitoring, which continuous thermodilution may relieve.
Does the Fourth-Generation FloTrac/Vigileo System accurately estimate cardiac output compared to continuous thermodilution in patients undergoing off-pump coronary artery bypass graft surgery?
Cohort (n=30)
Does the Fourth-Generation FloTrac/Vigileo System accurately estimate cardiac output compared to continuous thermodilution in patients undergoing off-pump coronary artery bypass graft surgery?
Mean Difference: -0.94 (95% CI -1.35–-0.52)
The Fourth-Generation FloTrac/Vigileo System showed clinically unacceptable precision and poor trending ability for cardiac output estimation during OPCAB surgery compared to continuous thermodilution.
Caution advised with fourth-generation FloTrac CO readings in OPCAB; leaves open need for prospective validation against thermodilution.
(1) Background: Previous studies reported limited performance of arterial pressure waveform-based cardiac output (CO) estimation (FloTrac/Vigileo system; CO-FloTrac) compared with the intermittent thermodilution technique (COint). However, errors due to bolus maneuver and intermittent measurements of COint could limit its use as a reference. The continuous thermodilution technique (COcont) may relieve such limitations. (2) Methods: The performance of CO-FloTrac was retrospectively assessed using continuous recordings of intraoperative physiological data acquired from patients who underwent off-pump coronary artery bypass graft (OPCAB) surgery with CO monitoring using both CO-FloTrac and COcont. Optimal time adjustments between the two measurements were determined based on R-squared values. (3) Results: A total of 134.2 h of data from 30 patients was included in the final analysis. The mean bias was –0.94 (95% CI, −1.35 to −0.52) L/min and the limits of agreements were −3.64 (95% CI, −4.44 to −3.08) L/min and 1.77 (95% CI, 1.21 to 2.57) L/min. The percentage error was 66.1% (95% CI, 52.4 to 85.8%). Depending on the time scale and the size of the exclusion zone, concordance rates ranged from 61.0% to 75.0%. (4) Conclusion: Despite the time adjustments, CO-FloTrac showed non-negligible overestimation, clinically unacceptable precision, and poor trending ability during OPCAB surgery.
No takes yet. Share an insight, caveat, or question.
Oh et al. (2022) conducted a cohort in Off-pump coronary artery bypass graft (OPCAB) surgery (n=30). Fourth-Generation FloTrac/Vigileo System (CO-FloTrac) vs. Continuous thermodilution technique (COcont) was evaluated on Cardiac output estimation performance (mean bias) (Mean bias -0.94 L/min, 95% CI -1.35 to -0.52). The fourth-generation FloTrac/Vigileo system showed non-negligible overestimation with a mean bias of -0.94 L/min (95% CI, -1.35 to -0.52) compared to continuous thermodilution during OPCAB surgery.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: