Key result
Fourth-generation FloTrac overestimates cardiac output by ~0.9 L/min with unacceptable precision versus continuous thermodilution.
Why the study?
Previous studies reported limited performance of FloTrac/Vigileo cardiac output estimation compared with intermittent thermodilution, but bolus errors and intermittent sampling limit intermittent thermodilution as a reference. Continuous thermodilution with time adjustment may overcome those reference limitations.
Does the fourth-generation FloTrac/Vigileo system provide accurate, precise, and interchangeable cardiac output estimation compared with time-adjusted continuous thermodilution during off-pump coronary artery bypass graft surgery?
Cohort (n=30)
No
Does the fourth-generation FloTrac/Vigileo system provide accurate, precise, and interchangeable cardiac output estimation compared with time-adjusted continuous thermodilution during off-pump coronary artery bypass graft surgery?
Even after accounting for the innate time delay of continuous thermodilution, the fourth-generation FloTrac/Vigileo system demonstrates clinically unacceptable precision (66.1% percentage error) and poor trending ability during OPCAB surgery.
Answers are grounded in this paper and cite their sources.
Caution advised with fourth-generation FloTrac cardiac output readings in OPCAB; confirms waveform analysis limitations and leaves open need for validated alternatives.
(1) Background: Previous studies reported limited performance of arterial pressure waveform-based cardiac output (CO) estimation (FloTrac/Vigileo system; CO-FloTrac) compared with the intermittent thermodilution technique (COint). However, errors due to bolus maneuver and intermittent measurements of COint could limit its use as a reference. The continuous thermodilution technique (COcont) may relieve such limitations. (2) Methods: The performance of CO-FloTrac was retrospectively assessed using continuous recordings of intraoperative physiological data acquired from patients who underwent off-pump coronary artery bypass graft (OPCAB) surgery with CO monitoring using both CO-FloTrac and COcont. Optimal time adjustments between the two measurements were determined based on R-squared values. (3) Results: A total of 134.2 h of data from 30 patients was included in the final analysis. The mean bias was –0.94 (95% CI, −1.35 to −0.52) L/min and the limits of agreements were −3.64 (95% CI, −4.44 to −3.08) L/min and 1.77 (95% CI, 1.21 to 2.57) L/min. The percentage error was 66.1% (95% CI, 52.4 to 85.8%). Depending on the time scale and the size of the exclusion zone, concordance rates ranged from 61.0% to 75.0%. (4) Conclusion: Despite the time adjustments, CO-FloTrac showed non-negligible overestimation, clinically unacceptable precision, and poor trending ability during OPCAB surgery.
No takes yet. Share an insight, caveat, or question.
Oh et al. (2022) conducted a cohort in Patients undergoing off-pump coronary artery bypass graft (OPCAB) surgery (n=30). Fourth-generation FloTrac/Vigileo system (CO-FloTrac) vs. Continuous thermodilution technique via pulmonary artery catheter (COcont) after individual time adjustment was evaluated on Mean bias (COcont minus CO-FloTrac) and precision (percentage error) of cardiac output estimation after individual time adjustment (MD -0.94 L/min (percentage error 66.1%), 95% CI -1.35 to -0.52). Fourth-generation FloTrac/Vigileo system overestimated cardiac output with a mean bias of -0.94 L/min and clinically unacceptable precision compared to continuous thermodilution.