Key result
Automated BP underestimates manual SBP by ~6 mm Hg in females using extra-large cuffs.
Why the study?
Automated blood pressure devices may be less accurate in females than males, but this required further investigation.
Does automated blood pressure measurement differ from manual measurement across sexes and cuff sizes in a general population?
Cross-Sectional (n=3,735)
Yes
Does automated blood pressure measurement differ from manual measurement across sexes and cuff sizes in a general population?
Automated oscillometric BP measurement may significantly underestimate systolic blood pressure compared to manual measurement in patients requiring larger cuff sizes, with a more pronounced effect in females.
May bias automated SBP readings with extra-large cuffs in practice; leaves open need for sex-specific validation in prospective hypertension studies.
BACKGROUND: Automated blood pressure (BP) devices may be less accurate in females than males, but this requires further investigation. This study aimed to determine sex differences in automated BP, measured with a single brand and model of device, compared with manual BP, with a focus on cuff sizes and associations with measures of adiposity. METHODS: Automated (Omron HEM-907XL) and manual BP were taken sequentially in a random order among a subsample of participants attending the US National Health and Nutrition Examination Survey, 2017 to 2018. Anthropometry and dual-energy x-ray absorptiometry were used to record body size and composition. Analyses, including multivariable regression to determine sex differences in BP, by cuff size, followed complex survey statistical principles. RESULTS: A total of 3735 participants (49.0% female [95% CI, 46.4-51.6], 45 years [43-46]) were included. In females, automated systolic BP (SBP) incrementally underestimated manual SBP across larger cuffs up to extra-large (-6.4 mm Hg [-8.0 to -4.9]). In males, automated SBP underestimated manual SBP only with extra-large cuffs (-2.4 mm Hg [95% CI-3.9 to -0.9]). Underestimation by automated SBP with extra-large cuffs was independently associated with all measures of body size indicative of increased adiposity in both females and males. Hypertension classification from automated and manual SBP had moderate agreement for adult/large cuffs (weighted kappa range 0.66-0.79) but weak agreement for extra-large cuffs (0.55-0.58) for females and males. CONCLUSIONS: The automated device used in this study underestimated manual SBP at larger cuff sizes, which was associated with indices of adiposity. Poorer accuracy of automated BP in larger cuff sizes could contribute to inequitable BP-related health care for females and males and requires further investigation.
No takes yet. Share an insight, caveat, or question.
Bui et al. (2025) reported a cross-sectional. Automated blood pressure measurement vs. Manual blood pressure measurement was evaluated on Difference between automated and manual systolic blood pressure using extra-large cuffs in females (MD -6.4 mm Hg, 95% CI -8.0 to -4.9). Automated blood pressure measurement underestimated manual systolic blood pressure at larger cuff sizes, particularly in females using extra-large cuffs (MD -6.4 mm Hg; 95% CI -8.0 to -4.9).