Purpose: To report the initial experience of patient-specific quality assurance (pQA) for the wobbling and line-scanning proton therapy at Samsung Medical Center. Materials and Methods:The pQA results of 89 wobbling treatments with 227 fields and 44 linescanning treatments with 118 fields were analyzed from December 2015 to June 2016.For the wobbling method, proton range and spread-out Bragg peak (SOBP) width were verified.For the line-scanning method, output and two-dimensional dose distribution at multiple depths were verified by gamma analysis with 3%/3 mm criterion.Results: The average range difference was -0.44 mm with a standard deviation (SD) of 1.64 mm and 0.1 mm with an SD of 0.53 mm for the small and middle wobbling radii, respectively.For the line-scanning method, the output difference was within ±3%.The gamma passing rates were over 95% with 3%/3 mm criterion for all depths.Conclusions: For the wobbling method, proton range and SOBP width were within the tolerance levels.For the line-scanning method, the output and two-dimensional dose distribution showed excellent agreement with the treatment plans.