Answers are grounded in this paper and cite their sources.
Given the limited access to healthcare resources, low-income settings require the development of affordable technology. Here we present the design and evaluation of a low-cost colorimeter applied to the non-invasive monitoring of Diabetes Mellitus through the detection of glucose in salival fluid. Samples were processed by the glucose oxidase-peroxidase enzymatic system and analyzed with the development equipment. A light emission diode of 532.5 nm was used as an excitation source and a RGB module was used as a receptor. A calibration curve to quantify the concentration of salivary glucose (0 to 18 mg/dL) was carried out by relating the RGB components registered with glucose concentrations, achieving a limit of detection of 0.17 mg/dL with a CV of 5% (n = 3). Salivary samples of diabetic and healthy volunteers were processed with the equipment showing an average concentration of 1.5519 ± 0.4511 mg/dL for the first and 4.0479 ± 1.6103 mg/dL for the last, allowing a discrimination between both groups. Results were validated against a UV-Vis-NIR spectrophotometer with a correspondence of R2 of 0.98194 between both instruments. Results suggest the potential application of the developed device to the sensitive detection of relevant analytes with a low-cost, user-friendly, low-power and portable instrumentation.
No takes yet. Share an insight, caveat, or question.
Dominguez et al. (2017) studied this question.