Key result
Large-scale pools show better commutability for PRA assays than small-scale or spiked pools.
Why the study?
The study was designed to evaluate the commutability of external quality assessment materials and candidate reference materials for plasma renin activity assays.
Large-scale pool external quality assessment materials demonstrate better commutability and are appropriate reference materials for plasma renin activity assays.
Large-scale EQA pools may improve PRA assay commutability; leaves open optimal material selection for clinical standardization.
OBJECTIVES: This study aims to evaluate the commutability of external quality assessment (EQA) materials and candidate reference materials (RMs) for plasma renin activity (PRA) assay. METHODS: Commutabilities of 16 candidate RMs were measured along with 40 clinical samples by the four different routine PRA assays, including three LC‒MS/MS assays and one chemiluminescence immunoassay. Sixteen candidate RMs included native/spiked human plasma pools (small-scale pools with <50 individuals) and current EQA materials (large-scale pools with >1,000 individuals). Difference in bias approach and linear regression with prediction interval approach were adopted to determine the commutability. Two-way variance analysis was used to estimate the effects of spiked and pool size on the commutability. Stability and homogeneity studies were performed. RESULTS: Precision and correlation performance of all assays was acceptable. In the difference in bias approach, the commutability results were not satisfactory (noncommutability: 14/16) and significant sample-specific effects were detected in assay pairs using different incubation buffers. For the prediction interval approach, no commutability was observed in the spiked small-scale pools; EQA materials (4/9) had more satisfactory commutability among all assays than the small-scale pools (2/7); RMs of large-scale pools tend to have better commutability no matter spiked or not. CONCLUSIONS: Commutable RMs were obtainable but challenging. Current EQA materials with relatively good commutability, stability, and homogeneity were appropriate RMs. Large-scale pools are tending to be commutable. Spiking in small-scale pools was not suggested to prepare RMs. MPs adopting a uniform incubation buffer would be preferable for further commutability research.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2023) studied Plasma renin activity measurement (n=56). Candidate reference materials (EQA materials and small-scale pools) was evaluated on Commutability determined by difference in bias and linear regression with prediction interval approaches. Current EQA materials (large-scale pools) demonstrated more satisfactory commutability (4/9) for plasma renin activity assays than small-scale pools (2/7), while spiked small-scale pools showed none.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: