The fluoride ion-selective electrode has been used in continuous flow analysis to determine the fluoride concentration in aqueous solution. The varying electrode potentials, E(t), were collected in the first 60 s by an interfacing computer, and the equilibrium potential of the electrode was predicted using an extrapolation method. This value agrees to within ±0.5 mV of the comparative value obtained by allowing sufficient time for the electrode to reach a steady state. Like flow-injection (FI), continuous flow (CF) analysis with this extrapolation method significantly shortens the time required for routine fluoride concentration measurements, but it also achieves a higher sensitivity than FI. The method has been applied to the determination of fluoride in city tap water.