This is the authors' abstract. We don't add key points for this paper.
An interface that enables automatic image acquisition during high-temperature experiments in an environmental SEM is developed. It is optimized to work on multiple regions of interest at multiple magnifications, performing image focusing (focus and astigmatism) and automatic re-centering of regions of interest. Its operation has been validated by monitoring two regions of interest of a nickel-based superalloy undergoing oxidation at 950 °C at different magnifications. Recording series of images at different magnifications on different regions of interest makes it possible to qualify the behavior of different areas of the sample in a single operation and/or to validate the reproducibility of the observations.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Lautru et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: