Key result
Combining MARD and PARD better characterizes CGM sensor performance.
Why the study?
A consistent evaluation of continuous glucose monitoring systems is lacking due to no consensus on reference methods or acceptance levels, leading to inconclusive performance assessments.
Combining MARD and PARD metrics provides a more robust evaluation of continuous glucose monitoring system accuracy and precision.
Answers are grounded in this paper and cite their sources.
Supports refined CGM accuracy assessment in research; prospective validation needed before clinical adoption.
BACKGROUND: Even though a Clinical and Laboratory Standards Institute proposal exists on the design of studies and performance criteria for continuous glucose monitoring (CGM) systems, it has not yet led to a consistent evaluation of different systems, as no consensus has been reached on the reference method to evaluate them or on acceptance levels. As a consequence, performance assessment of CGM systems tends to be inconclusive, and a comparison of the outcome of different studies is difficult. MATERIALS AND METHODS: Published information and available data (as presented in this issue of Journal of Diabetes Science and Technology by Freckmann and coauthors) are used to assess the suitability of several frequently used methods [International Organization for Standardization, continuous glucose error grid analysis, mean absolute relative deviation (MARD), precision absolute relative deviation (PARD)] when assessing performance of CGM systems in terms of accuracy and precision. RESULTS: The combined use of MARD and PARD seems to allow for better characterization of sensor performance. The use of different quantities for calibration and evaluation, e.g., capillary blood using a blood glucose (BG) meter versus venous blood using a laboratory measurement, introduces an additional error source. Using BG values measured in more or less large intervals as the only reference leads to a significant loss of information in comparison with the continuous sensor signal and possibly to an erroneous estimation of sensor performance during swings. Both can be improved using data from two identical CGM sensors worn by the same patient in parallel. CONCLUSIONS: Evaluation of CGM performance studies should follow an identical study design, including sufficient swings in glycemia. At least a part of the study participants should wear two identical CGM sensors in parallel. All data available should be used for evaluation, both by MARD and PARD, a good PARD value being a precondition to trust a good MARD value. Results should be analyzed and presented separately for clinically different categories, e.g., hypoglycemia, exercise, or night and day.
No takes yet. Share an insight, caveat, or question.
Obermaier et al. (2013) conducted a review in Diabetes. Mean absolute relative deviation (MARD) and precision absolute relative deviation (PARD) was evaluated on Performance of CGM systems in terms of accuracy and precision. The combined use of mean absolute relative deviation (MARD) and precision absolute relative deviation (PARD) allows for better characterization of continuous glucose monitoring sensor performance.
Synapse has enriched 10 related papers. Consider them for comparative context: