Answers are grounded in this paper and cite their sources.
Isotopic analyses of plutonium and low-enriched uranium mixtures with particle sizes of 0.6–3.3 μm were performed using thermal ionization mass spectrometry with a continuous heating method to verify its effectiveness for the accurate analysis of minor isotopes without sample pretreatment. The mixed particles used in this study were prepared from a mixed solution of plutonium (SRM 947) and uranium (U010, 235 U 1% enriched) reference materials. The isotope ratios for plutonium in the individual mixed particles, including 238 Pu/ 239 Pu, 241 Pu/ 239 Pu as well as 240 Pu/ 239 Pu, and 242 Pu/ 239 Pu, were in good agreement with the certified values despite the isobaric interference of 238 U and 241 Am. The isotope ratios for uranium in the mixed particles also agreed well with the certified values within the range of error. However, the isotope ratios for minor isotopes, such as 234 U and 236 U, in the particles with diameters of less than approximately 1.8 μm could not be measured because numbers of 234 U and 236 U atoms in analyzed particles are too low. These results indicate that thermal ionization mass spectrometry with a continuous heating method is applicable for the analysis of trace amounts of plutonium isotopes, including 238 Pu and 241 Pu, despite the presence of the respective isobars 238 U and 241 Am in the microsamples.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2017) studied this question.