Answers are grounded in this paper and cite their sources.
The voltammetric behavior of paraquat was investigated at hydroxyapatite‐modified carbon paste electrode HAP‐CPE in K2SO4. A method was developed for the detection of the trace of this herbicide, based on their redox reaction. The reduction peaks of paraquat were observed around −0.70 V and −1.00 V (vs. SCE) in square‐wave voltammetry. Experimental conditions were optimized by varying the accumulation time, apatite loading and measuring solution pH. Calibration plots were linear under the optimized parameters over the herbicide's concentration range 8–200×10−7 mol L−1, with a detection and quantification limits about 1.5×10−8 mol L−1 and 6.4 10−8 mol L−1, respectively.