Key result
Novel wearable wireless sphygmomanometer estimates BP within <5 mmHg of reference equipment.
Why the study?
Traditional ambulatory blood pressure monitors require long and obtrusive tubing interconnects between the device and the cuff, prompting the design of an unobtrusive, wireless wearable monitoring system.
Does a wearable wireless-enabled oscillometric sphygmomanometer accurately estimate blood pressure compared to reference equipment in human subjects?
Does a wearable wireless-enabled oscillometric sphygmomanometer accurately estimate blood pressure compared to reference equipment in human subjects?
A novel wearable wireless-enabled oscillometric sphygmomanometer demonstrates accurate blood pressure estimation with a smaller and lighter form factor than traditional ambulatory monitors.
May support wearable BP monitoring development; leaves open larger ambulatory validation before clinical use.
This article presents the design of an unobtrusive and wireless-enabled blood pressure (BP) monitoring system that is suitable for ambulatory use. By adopting low-profile electromechanical actuators and a compact printed circuit board design, this lightweight device can be worn directly on the occlusive cuff, therefore eliminating the need of a long and obtrusive tubing interconnect between the device and the cuff, as seen in traditional ambulatory BP monitors (ABPM). Instead of executing the BP estimation algorithm directly on the device, the proposed design rather sends the raw oscillometric signal through a Bluetooth Low Energy link, thus granting any Bluetooth-enabled device to gather and process the signal using a dedicated application. This in turn allows to assess several BP estimation algorithms found in the literature without being limited by the device resources. Three of them were tested with the designed prototype and validated with a reference equipment on 11 subjects. Overall, two of the algorithms revealed a mean absolute difference with the reference equipment of less than 5 mmHg and almost zero bias along with a standard deviation of less than 6 mmHg. Reproducibility results shown a mean difference between successive measurements of less than 3.1 mmHg and a standard deviation of less than 2.4 mmHg. The assembled prototype dimensions are 63.8 × 134.8 × 24.8 mm and features an autonomy of 63.1 hours. Comparison with commercial ABPM devices shown that the proposed design is 18% to 33% smaller volume-wise, 5% to 27% weight-wise and height is reduced by 17% to 25%.
No takes yet. Share an insight, caveat, or question.
Juteau et al. (2020) studied Blood pressure estimation (n=11). Wearable wireless-enabled oscillometric sphygmomanometer vs. Reference equipment was evaluated on Mean absolute difference and standard deviation compared to reference equipment. A novel wearable wireless-enabled oscillometric sphygmomanometer demonstrated a mean absolute difference of less than 5 mmHg and standard deviation of less than 6 mmHg compared to reference equipment.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: