Key result
Transcatheter perimembranous VSD closure using Amplatzer Occluder 2 achieves 95% implant success without procedural complications.
Why the study?
Transcatheter closure of perimembranous ventricular septal defects has been associated with substantial risk of complete heart block, limiting its use.
Does transcatheter closure using the Amplatzer Membranous VSD Occluder 2 safely and effectively treat perimembranous ventricular septal defects without causing complete heart block?
Cohort (n=19)
Yes
Does transcatheter closure using the Amplatzer Membranous VSD Occluder 2 safely and effectively treat perimembranous ventricular septal defects without causing complete heart block?
The Amplatzer Membranous VSD Occluder 2 appears feasible and safe for transcatheter closure of perimembranous VSDs, notably avoiding the complication of complete heart block in this initial cohort.
Answers are grounded in this paper and cite their sources.
May extend transcatheter options for perimembranous VSD; hypothesis-generating pending randomized evidence.
OBJECTIVES: To describe the initial world experience and mid-term follow-up of perimembranous ventricular septal defect (pmVSD) closure with a newly designed occluder. BACKGROUND: Transcatheter closure of pmVSDs has been associated with a substantial risk of complete heart block, prompting many centers to abandon this intervention. METHODS: A prospective multicenter cohort study was conducted on patients with pmVSD undergoing catheter closure using the Amplatzer(®) Membranous VSD Occluder 2 in the initial 4 pilot centers. RESULTS: Nineteen patients, median age 6 years (range 1.4-62 years), were enrolled and followed for 14 ± 3 months (range 8-20 months). The median weight was 26 kg (range 9.3-96 kg) and the mean Qp/Qs ratio was 1.8 ± 0.7. The defect on left ventricular side measured 9.9 ± 3.5 mm and the orifice on right ventricular side was 8.1 ± 2.8 mm by echocardiography. Mean device size was 9.4 ± 2.4 mm (range 5-14 mm). An eccentric device was employed in 9 patients (47%) and a concentric device in 10 (53%). Overall, 18 patients (95%) had successful device implants. Procedural time was 122 ± 39 min. There were no procedural complications. Mild residual shunting was initially observed in 14 (78%) patients. At last follow-up, mild residual shunting persisted in only 3 (17%) patients. There was no significant increase in aortic or tricuspid regurgitation. No patient had any degree of AV block, although one developed a transient left anterior fascicular block. Holter evaluation, obtained in all patients, was unremarkable in all. CONCLUSIONS: This early cohort experience using a novel adapted transcatheter closure device for pmVSD suggests that the procedure is feasible, safe, and effective.
No takes yet. Share an insight, caveat, or question.
Tzikas et al. (2013) conducted a cohort in perimembranous ventricular septal defect (pmVSD) (n=19). Amplatzer Membranous VSD Occluder 2 was evaluated on Successful device implants. Transcatheter closure of perimembranous ventricular septal defects using the Amplatzer Membranous VSD Occluder 2 achieved successful implantation in 95% of patients with no procedural complications.
Synapse has enriched 10 related papers. Consider them for comparative context: