Key result
Smartwatch PPG data analyzability averages ~72% daily and correlates inversely with physical activity.
Why the study?
Continuous heart rate monitoring via PPG-based mobile health technologies is promising for early detection of sustained cardiac arrhythmias but is impaired by motion artifacts.
Observational (n=41)
Smartwatch-derived PPG data analyzability is strongly inversely related to physical activity, suggesting that continuous arrhythmia screening is most effective during sleep.
Answers are grounded in this paper and cite their sources.
Smartwatch PPG data analyzability declines with rising activity; leaves open reliable arrhythmia detection in daily life and needs prospective validation.
BACKGROUND Continuous heart rate monitoring via mobile health technologies based on photoplethysmography (PPG) has great potential for the early detection of sustained cardiac arrhythmias such as atrial fibrillation. However, PPG measurements are impaired by motion artifacts. OBJECTIVE The aim of this investigation was to evaluate the analyzability of smartwatch-derived PPG data during everyday life and to determine the relationship between the analyzability of the data and the activity level of the participant. METHODS A total of 41 (19 female and 22 male) adults in good cardiovascular health (aged 19-79 years) continuously wore a smartwatch equipped with a PPG sensor and a 3D accelerometer (Cardio Watch 287, Corsano Health BV) for a period of 24 hours that represented their individual daily routine. For each participant, smartwatch data were analyzed on a 1-minute basis by an algorithm designed for heart rhythm analysis (Preventicus Heartbeats, Preventicus GmbH). As outcomes, the percentage of analyzable data (PAD) and the mean acceleration (ACC) were calculated. To map changes of the ACC and PAD over the course of one day, the 24-hour period was divided into 8 subintervals comprising 3 hours each. RESULTS Univariate analysis of variance showed a large effect (ηp2> 0.6; P<.001) of time interval (phase) on the ACC and PAD. The PAD ranged between 34% and 100%, with an average of 71.5% for the whole day, which is equivalent to a period of 17.2 hours. Between midnight and 6 AM, the mean values were the highest for the PAD (>94%) and the lowest for the ACC (<6×10-3 m/s2). Regardless of the time of the day, the correlation between the PAD and ACC was strong (r=–0.64). A linear regression analysis for the averaged data resulted in an almost perfect coefficient of determination (r2=0.99). CONCLUSIONS This study showed a large relationship between the activity level and the analyzability of smartwatch-derived PPG data. Given the high yield of analyzable data during the nighttime, continuous arrhythmia screening seems particularly effective during sleep phases.
No takes yet. Share an insight, caveat, or question.
Merschel et al. (2021) conducted an observational in good cardiovascular health (n=41). Smartwatch-derived PPG data monitoring was evaluated on percentage of analyzable data (PAD) and mean acceleration (ACC) (r=-0.64, p=<.001). Smartwatch-derived PPG data analyzability averaged 71.5% over 24 hours and was strongly inversely correlated with activity level (r=-0.64), with the highest analyzability (>94%) during nighttime.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: