Answers are grounded in this paper and cite their sources.
BACKGROUND AND OBJECTIVES: HydroMARK® is a newly available biopsy marker for image-guided needle biopsies of non-palpable breast lesions. Objective was to determine if the marker could be utilized independently for lesion localization using intra-operative ultrasound alone. METHODS: A single institution retrospective review identified patients who underwent surgical excision of breast lesions after placement of the HydroMARK®. Endpoints included intra-operative visualization of the marker, successful excision of the lesion, and presence of the marker on specimen radiograph. RESULTS: The study included 31 lesions in 25 patients. Twenty-nine (93.6%) HydroMARKSs® were adequately visualized by intra-operative ultrasound. Intra-operative ultrasound without pre-operative placement of a localizing device was successful for localization in six cases (19.4%). Intra-operative difficulties were encountered in 16 of 31 (51.6%) procedures. This included either extrusion of the marker when the biopsy tract was transected in 14 (45.2%) cases or migration of the marker prior to the procedure in two (6.4%) cases. The marker was visualized on specimen radiograph in 15 (48.4%) cases. CONCLUSIONS: While intraoperative sonographic visibility was excellent, a large number of excisions were associated with extrusion of the marker. Modifications are needed to improve acceptability of this marker for intra-operative localization independent of pre-operative wire or seed localization.
No takes yet. Share an insight, caveat, or question.
Klein et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: