Key result
Sensitive HPLC assay measures plasma normetanephrine and metanephrine up to 2000 pg without acetaminophen interference.
Why the study?
A method to measure unconjugated plasma normetanephrine and metanephrine without interference from acetaminophen was needed due to its widespread use.
A sensitive HPLC assay for plasma metanephrines was developed that avoids interference from acetaminophen, facilitating accurate biochemical diagnosis of pheochromocytoma.
Answers are grounded in this paper and cite their sources.
May enable reliable pheochromocytoma screening despite acetaminophen; leaves open prospective validation before clinical use.
BACKGROUND: Pheochromocytoma is a rare cause of hypertension resulting from increased catecholamine secretion. We aimed to develop a method to measure unconjugated plasma normetanephrine (NMN) and metanephrine (MN) without interference from acetaminophen, a widely prescribed drug for headaches. METHODS: Plasma samples were obtained from 48 subjects (23 males, 25 females; mean age, 49 +/- 14 years; hypertension, n = 37) under resting conditions. Following extraction on solid-phase cation-exchange columns, unconjugated metanephrines were analyzed by HPLC with electrochemical detection and with 4-hydroxy-3-methoxybenzylamine as an internal standard. Catecholamines were measured by HPLC. RESULTS: The assays were linear up to 2000 pg for NMN and for MN. Intraassay imprecisions (CVs) were 4.7% for NMN and 7.0% for MN, and the interassay CV was 12% for both NMN and MN. The limit of detection was 11 fmol for NMN and 17 fmol for MN. Ingestion of acetaminophen or its addition to plasma did not interfere with the MN peaks. Plasma NMN and MN were positively correlated (r = 0.52 and 0.49, respectively; P <0.01 for both) with the respective catecholamines. Plasma NMN (r = 0.27; P = 0.02) but not MN positively correlated with age, whereas only plasma catecholamines (and not metanephrines) were positively correlated (P <0.05) with diastolic blood pressure. CONCLUSIONS: This sensitive MN assay is not affected by simultaneous acetaminophen medication, and reveals a correlation of metanephrines with plasma and urinary catecholamines and age but not with blood pressure.
No takes yet. Share an insight, caveat, or question.
Roden et al. (2001) studied Hypertension (n=48). HPLC assay for unconjugated plasma metanephrines was evaluated on Assay linearity, precision, and interference by acetaminophen. A sensitive HPLC assay for plasma normetanephrine and metanephrine was developed that is linear up to 2000 pg and unaffected by simultaneous acetaminophen medication.
Synapse has enriched 10 related papers. Consider them for comparative context: