Key result
Long-shaft water testing in minimally invasive mitral repair demonstrates ~100% MR ≤1 without major adverse events.
Why the study?
Reproducible water testing with less air contamination during minimally invasive mitral valve repair may help achieve successful repair and prevent air embolism.
Does water testing with a long shaft irrigation device improve assessment of valve competence and prevent air embolism in patients undergoing minimally invasive mitral valve repair?
Observational (n=104)
Does water testing with a long shaft irrigation device improve assessment of valve competence and prevent air embolism in patients undergoing minimally invasive mitral valve repair?
A novel water testing method using a long shaft irrigation device during minimally invasive mitral valve repair safely confirms valve competence while minimizing air introduction.
Answers are grounded in this paper and cite their sources.
May facilitate reproducible water testing in minimally invasive mitral repair; hypothesis-generating, needs prospective validation.
Minimally invasive mitral valve repair has become standard. We hypothesized that reproducible water testing with less air contaminations may help achieve successful repair and prevent air embolism. We used a long shaft irrigation device connected to an infuser pump to water test mitral valve repair in 104 consecutive patients with severe mitral regurgitation due to type II mitral valve dysfunction. All patients underwent mitral valve repair via small right thoracotomy. Water testing was performed by simultaneously monitoring root pressure to confirm valve competence and adequate leaflet coaptation with suitable pressure. Irrigation was stopped when root pressure reached 60 mm Hg. Simultaneous transesophageal echocardiography showed a negligible amount of air introduction. Predischarge echocardiography revealed residual mitral regurgitation of less than or equal to 1 in all patients, and no patient experienced any major adverse event such as stroke or perioperative myocardial infarction. This novel water test allowed us to correctly check mitral valve competence and reduce air contamination.
No takes yet. Share an insight, caveat, or question.
Fujita et al. (2017) conducted an observational in Severe mitral regurgitation due to type II mitral valve dysfunction (n=104). Water testing using a long shaft irrigation device connected to an infuser pump was evaluated on Residual mitral regurgitation and major adverse events (stroke or perioperative myocardial infarction). Water testing with a long shaft irrigation device during minimally invasive mitral valve repair resulted in residual mitral regurgitation ≤1 and no major adverse events in all 104 patients.
Synapse has enriched 10 related papers. Consider them for comparative context: