The authors report on the stability and complications of 73 LICOX brain ti-pO2-microcatheters in 70 patients. Mean monitoring time was 7.5 +/- 4.0 days. Patients prone to cerebral hypoxia (after severe head injury (GCS < 9) or a subarachnoid hemorrhage) had a ti-pO2-microcatheter inserted next to the ICP-probe in the typical frontal position. After the first 15 insertions, instead of the 3-way-screw (needing a 6 mm burrhole), a 1-way-screw (needing a 2.7 mm burrhole) was used for fixation in the bone; by doing so, the procedure can be performed in the ICU and takes only 15 min. Whenever possible a calibration at room air (to determine the sensitivity-drift) and in oxygen free solution (to determine the zero-drift) was performed after removal of the catheters. Ideally the expected pO2 at room air was around 154 mmHg (temperature dependent) and at zero calibration 0 mmHg. Mean sensitivity-drift for 54 catheters was -8.5 +/- 15.4%. Dividing the catheters into groups, depending on the duration of monitoring (1-4, 5-8 and 9-16 days), revealed that the greatest part of the (negative) sensitivity-drift occurred during day 1-4 after insertion. After 1 week of monitoring sometimes a positive drift occurred (being far less than the negative drift during the first 4 days). Compared to the old catheters (-10.3 +/- 17.3%) (on the first half of the patients) the new ones showed a lower sensitivity-drift (-6.8 +/- 13.4%). The zero-drift of 56 catheters was low with mean drift after 7.5 +/- 4.0 days of 1.5 +/- 1.5 mmHg. Here also the highest drift occurred on day 1-4 after insertion. No infection was seen and 2 times (2.7%) a small hematoma, not needing evacuation occurred. As the ti-pO2-catheter (having a smaller diameter) and the ICP-catheter were inserted at the same time, one cannot distinguish which catheter caused the hematoma. A possible explanation for the occurrence of the two hematomas is the insertion of the catheters too close to the midline. The authors conclude that LICOX ti-pO2-monitoring is a safe and reliable method. Further decrease of the complication rate and increase of the catheter-stability may be expected.
No takes yet. Share an insight, caveat, or question.
Dings et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: