Key result
Device-reported and calculated oscillometric MAP show good agreement but correlate poorly with manual measurements.
Why the study?
Does oscillometric MAP measurement agree with manual MAP measurement in children in the emergency department?
Observational (n=347)
No
Does oscillometric MAP measurement agree with manual MAP measurement in children in the emergency department?
Effect estimate: ICC 0.83 (95% CI 0.79-0.86)
In pediatric emergency department patients, calculated and device-reported oscillometric MAP measurements show good agreement, whereas manual MAP measurements correlate poorly with oscillometric methods.
Supports preferring oscillometric over manual MAP in pediatric emergencies; leaves open outcome impact and practice change.
OBJECTIVE: To evaluate agreement among three approaches to mean arterial pressure (MAP) measurement in children. METHODS: We performed a single-center prospective study of a convenience sample of children (<18 y) presenting to a pediatric emergency department. We compared MAP agreement using the intraclass correlation coefficient (ICC) measured in three ways: calculated via auscultation, calculated from oscillometric systolic and diastolic measurements, and measurement via oscillometry. RESULTS: We included 347 patients (median age 8 y, IQR: 4 to 13). The highest ICC was observed between calculated and device-reported oscillometric MAP measurements (0.83, 95% CI: 0.79-0.86), indicating good agreement. Agreement was poor between manual and device-reported oscillometric MAP (ICC: 0.33, 95% CI: 0.22-0.43) and between manual and calculated oscillometric MAP (ICC: 0.38, 95% CI: 0.20-0.52). The median difference between calculated oscillometric and device-reported oscillometric MAP was 1.0 mm Hg (IQR: 0.0 to 2.3), compared with -3.3 mm Hg (IQR: -8.3 to 2.0) between manual and device-reported oscillometric MAP and -4.0 mm Hg (IQR: -9.3 to 0.3) between manual and calculated oscillometric MAP. Bland-Altman plots suggested that manual MAP measurements were slightly lower than both device-reported oscillometric and calculated oscillometric MAP measurements. Agreement between calculated oscillometric and device-reported oscillometric MAP was highest among children <2 years and lowest among children 12 to <18 years. Agreement involving manual MAP remained poor across most age groups. Similar patterns were observed when stratified by weight and patient state. CONCLUSIONS: Calculated oscillometric MAP had good agreement with device-reported oscillometric MAP, while manual MAP had poorer agreement with either oscillometric measure. Further study is needed to determine the clinical implications of these differences across age, weight, and patient-state subgroups.
No takes yet. Share an insight, caveat, or question.
Diasso-Adamou et al. (2026) conducted an observational in Children presenting to a pediatric emergency department (n=347). Calculated oscillometric MAP vs. Device-reported oscillometric MAP and manual MAP was evaluated on Agreement among three approaches to mean arterial pressure (MAP) measurement using intraclass correlation coefficient (ICC) (ICC 0.83, 95% CI 0.79-0.86). Calculated oscillometric MAP had good agreement with device-reported oscillometric MAP (ICC 0.83; 95% CI 0.79-0.86), whereas manual MAP had poor agreement with either oscillometric measure.