Key result
Transcatheter VSD closure with the Nit-Occlud Lê coil achieves ~99% procedural success.
Why the study?
Transcatheter VSD closure has achieved encouraging results but more follow-up studies are needed to assess safety and efficacy.
Does transcatheter closure using the Nit-Occlud Lê VSD Coil safely and effectively treat ventricular septal defects in pediatric patients?
Observational (n=80)
No
Does transcatheter closure using the Nit-Occlud Lê VSD Coil safely and effectively treat ventricular septal defects in pediatric patients?
The Nit-Occlud Lê VSD Coil provides a safe and effective transcatheter closure option for various VSD morphologies in pediatric patients, achieving a 97.5% complete occlusion rate at 1 year.
Answers are grounded in this paper and cite their sources.
Should not change VSD closure practice; leaves open need for randomized confirmation of long-term outcomes.
OBJECTIVES: To assess safety, efficacy and follow-up results of transcatheter closure of ventricular septal defect (VSD) using Nit-Occlud_ Lê VSD Coil (pfm medical, KÖln, Germany). BACKGROUND: Transcatheter VSD closure has achieved encouraging results but more follow-up studies are needed. PATIENTS AND METHODS: Between January 2012 and December 2013 in the cardiology department, Tanta University Hospital, Tanta, Egypt, 80 patients underwent percutaneous VSD closure using Nit-Occlud_ Lê VSD Coil. Early and mid- term follow-up was done for 3 years, follow-up was concluded in 2016. RESULTS: The mean age of patients was 5.34 ± 3 years, and their mean weight was 17.24 ± 8.17 kg. Overall, 77 of 80 patients had perimembranous VSD with aneurysmal tissue; eight had multiple right ventricular exits, 14 had deficient aortic rim, two had high outlet muscular, and one had Gerbode defect. The procedure was successful in 98.75% of patients, and was aborted in one patient because of the development of complete heart block and the coil had to be removed. The mean procedure time was 104.98 ± 9.50 minutes. The mean fluoroscopy time was 30.58 ± 2.79 minutes. The immediate complete occlusion rate was 62%, which increased to 82.3% on the second day, and 94.9% by the 3rd month, and 97.5% by 1 year. There was a significant decrease in mitral incompetence after 6 months of follow-up (p = 0.002), and only one patient had trivial aortic incompetence prior to the procedure that remained the same during follow-up period. CONCLUSION: Using Nit-Occlud_ Lê VSD-Coil to close VSD is safe and feasible in VSDs with various morphology.
No takes yet. Share an insight, caveat, or question.
Shedoudy et al. (2018) conducted an observational in Ventricular septal defect (n=80). Transcatheter closure using Nit-Occlud Lê VSD Coil was evaluated on Procedural success. Transcatheter closure of ventricular septal defects using the Nit-Occlud Lê VSD Coil achieved a 98.75% procedural success rate and a 97.5% complete occlusion rate at 1 year.
Synapse has enriched 10 related papers. Consider them for comparative context: