The construction, performance characteristics, and application of a polymeric membrane coated on a graphite electrode with unique selectivity toward histidine are reported. The electrode was prepared by incorporating chloro(5,10,15,20-tetraphenylporphyrinato) manganese(III) [Mn(TPP)Cl] into a plasticized poly(vinyl chloride) membrane. The influences of membrane composition, pH, and foreign ions were investigated. Calibration plots with near Nernstian slopes for histidine were observed, -55.4 mV/decade, over a linear range of four decades of concentration (1 × 10(-)(5) to 1 × 10(-)(1) M). The electrode has a detection limit of 5 × 10(-)(6) M histidine and shows a fast response time of about 1 min. The electrode shows high selectivities toward histidine over several amino acids and common inorganic anions.
No takes yet. Share an insight, caveat, or question.
Amini et al. (1999) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: