This is the authors' abstract. We don't add key points for this paper.
The hepatic peptide hormone hepcidin plays a central role in body iron metabolism. Despite its promise as a biomarker, the availability of high-sensitive hepcidin assays is still limited. We developed and validated a RadioImmunoAssay (RIA) to measure hepcidin quantitatively in human serum. This assay exhibited a very low detection limit (0.02 microg/l), low imprecision (coefficient of variation-range 4.4-6.2%) and good linearity and recovery (range: 81-105%). Hepcidin levels of samples of controls and patients with iron deficiency and inflammation showed an excellent correlation with our previously described quantitative time-of-flight mass spectrometry assay (range 2.5-266.8 microg/l, r = 0.92, P < 0.0001). The RIA detected: (i) differences in mean hepcidin levels between men (n = 29) and women (n = 35), (ii) differences between individuals of different HFE-genotypes (n = 60) and (iii) daily increases in hepcidin levels (n = 64). The assay (i) is easy to perform and many samples can be processed within one assay-run, (ii) shows accurate, reproducible and high-sensitive measurements and (iii) is anticipated to be particularly useful to study the effects of pathological and physiological stimuli on hepcidin levels in the lower range.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Grebenchtchikov et al. (2009) studied this question.