Cadmium-zinc-telluride myocardial perfusion imaging yielded nondiagnostic image quality in 81% of patients with BMI ≥40 kg/m2, with a BMI cutoff of 39 kg/m2 for diagnostic quality.
Why the study?
The impact of increased body mass on myocardial perfusion imaging quality using a cadmium-zinc-telluride γ-camera was evaluated due to concerns about image quality in obese patients.
Does cadmium-zinc-telluride myocardial perfusion imaging provide adequate diagnostic image quality in obese patients compared to conventional SPECT?
Observational (n=81)
Does cadmium-zinc-telluride myocardial perfusion imaging provide adequate diagnostic image quality in obese patients compared to conventional SPECT?
Answers are grounded in this paper and cite their sources.
Patients with a BMI ≥ 40 kg/m2 have a high rate of nondiagnostic myocardial perfusion images with CZT cameras and should be scheduled for conventional SPECT imaging.
UNLABELLED: We have evaluated the impact of increased body mass on the quality of myocardial perfusion imaging using a latest-generation γ-camera with cadmium-zinc-telluride semiconductor detectors in patients with high (≥40 kg/m(2)) or very high (≥45 kg/m(2)) body mass index (BMI). METHODS: We enrolled 81 patients, including 18 with no obesity (BMI < 30 kg/m(2)), 17 in World Health Organization obese class I (BMI, 30-34.9 kg/m(2)), 15 in class II (BMI, 35-39.9 kg/m(2)), and 31 in class III (BMI ≥ 40 kg/m(2)), including 15 with BMI ≥ 45 kg/m(2). Image quality was scored as poor (1), moderate (2), good (3), or excellent (4). Patients with BMI ≥ 45 kg/m(2) and nondiagnostic image quality (≤2) were rescanned after repositioning to better center the heart in the field of view. Receiver-operating-curve analysis was applied to determine the BMI cutoff required to obtain diagnostic image quality (≥3). RESULTS: Receiver-operating-curve analysis resulted in a cutoff BMI of 39 kg/m(2) (P < 0.001) for diagnostic image quality. In patients with BMI ≥ 40 kg/m(2), image quality was nondiagnostic in 81%; after CT-based attenuation correction this decreased to 55%. Repositioning further improved image quality. Rescanning on a conventional SPECT camera resulted in diagnostic image quality in all patients with BMI ≥ 45 kg/m(2). CONCLUSION: Patients with BMI ≥ 40 kg/m(2) should be scheduled for myocardial perfusion imaging on a conventional SPECT camera, as it is difficult to obtain diagnostic image quality on a cadmium-zinc-telluride camera.
No takes yet. Share an insight, caveat, or question.
Fiechter et al. (2012) conducted an observational in Obesity (n=81). Cadmium-zinc-telluride myocardial perfusion imaging vs. Conventional SPECT camera was evaluated on Diagnostic image quality (score ≥3) (p=<0.001). Cadmium-zinc-telluride myocardial perfusion imaging yielded nondiagnostic image quality in 81% of patients with BMI ≥40 kg/m2, with a BMI cutoff of 39 kg/m2 for diagnostic quality.