It is highly demanding to identify healthy and non-healthy species in a heterogeneous environment such as human tissues. In such a case, one identifier, such as a spectral fingerprint, might be inadequate. Therefore, additional identification is required, for instance, a polarisation measurement. In view of that, the development of a spectropolarimeter that captures two cross-polarised arrays of spectral images is a key requirement. To meet this requirement, an imager optical setup has been designed to provide spatial, spectral and polarisation preference information for species that exist in a heterogeneous environment, such as in medical tissue samples. The spectral and polarisation information is obtained employing an acousto-optic tunable filter and a polarising beam splitter, respectively. The optical imager is designed to operate in the visible-near infrared range (450–850 nm) with a spectral resolution of 3 nm. The spectropolarimeter design along with optical characterisation results are reported.
No takes yet. Share an insight, caveat, or question.
Abdlaty et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: