Answers are grounded in this paper and cite their sources.
BACKGROUND: Robotic end-effectors for single port endoscopic surgery (SPS) require a manual change of vision field that slows surgery and increases the degrees of freedom (DOFs) of the manipulator. METHODS: A new surgical prototype robot has dynamic vision field control and a master controller to manipulate the endoscopic view. It uses positioning (4 DOF) and sheath (2 DOF) manipulators for vision field control, and dual tool tissue manipulators (gripping, 5 DOF; cautery, 3 DOF). RESULTS: The robot is feasible in vitro. 'Cut and vision field control' (using tool manipulators) was suitable for precise cutting tasks in risky areas; 'cut by vision field control' (using the vision field control manipulator) was effective for rapid macro cutting of tissues. A resection was performed using a combination of both methods. CONCLUSIONS: The novel robotic system is feasible, but further studies are needed to address its performance in vivo.
No takes yet. Share an insight, caveat, or question.
Kobayashi et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: