A novel wireless sensor system combining ECG and EDA for stress detection demonstrates high technical performance with negligible transmission delays.
This paper describes the design of a two channels electrodermal activity (EDA) sensor and two channels electrocardiogram (ECG) sensor. The EDA sensors acquire data on the hands and transmit them to the ECG sensor with wireless WiFi communication for increased wearability. The sensors system acquires two EDA channels to improve the removal of motion artifacts that take place if EDA is measured on individuals who need to move their hands in their activities. The ECG channels are acquired on the chest and the ECG sensor is responsible for aligning the two ECG traces with the received packets from EDA sensors; the ECG sensor sends via WiFi the aligned packets to a laptop for real time plot and data storage. The metrological characterization showed high-level performances in terms of linearity and jitter; the delays introduced by the wireless transmission from EDA to ECG sensor have been proved to be negligible for the present application.
No takes yet. Share an insight, caveat, or question.
Antonio Affanni (2020) studied Stress. Wireless Sensors System (ECG and EDA) was evaluated on Metrological characterization (linearity and jitter). The designed wireless sensor system for stress detection using ECG and EDA showed high-level performances in terms of linearity and jitter, with negligible wireless transmission delays.