Key result
Valved stent delivery systems cause significant intimal damage across all aortic segments regardless of catheter.
Why the study?
The extent of intimal damage to the human aorta after deployment of different conventional valved stent delivery systems during transcatheter valve implantation was not well characterized.
Does the passage of transcatheter valved stent delivery systems cause intimal damage in human aortas in vitro?
Does the passage of transcatheter valved stent delivery systems cause intimal damage in human aortas in vitro?
Transfemoral passage of transcatheter valve delivery systems causes significant intimal damage in human aortas in vitro, highlighting the need for lower profile and more flexible devices to avoid late aortic complications.
Significant intimal damage from TAVR delivery systems may signal late aortic risks; hypothesis-generating for lower-profile devices.
Transcatheter heart valve implantation can be performed transapically and transfemorally. The transfemoral way to the aortic valve is significantly longer than the transapical one. The aim of this study was to analyse the intima of 15 human aortas after the deployment of different conventional valved stent delivery systems. Fifteen human aortas have been analysed (77 ± 8.4 years). These aortas were preserved with formalin and explanted from the common iliac arteries to the ascending aorta. After protocolling all relevant vascular parameters, the deployment force of different conventional valved stent delivery systems was analysed. After that, the intima was closely investigated by endoscopy. The deployment force of the old catheter was not different from the actual system. The endoscopic investigation revealed significant intimal damages in all parts of the aorta after deployment of the delivery system. This study demonstrated that the passage to the aortic valve can result in significant intimal damage regardless of the used deployment catheter. Efforts are necessary to lower the profile of the deployment devices to increase their flexibility. The intima of the aorta and possible damage have to get back into the physicians' focus to avoid possible late aortic complications.
No takes yet. Share an insight, caveat, or question.
Heinisch et al. (2012) studied Transcatheter heart valve implantation (n=15). Deployment of conventional valved stent delivery systems vs. Old catheter vs actual system was evaluated on Intimal damage and deployment force. Deployment of valved stent delivery systems in human aortas resulted in significant intimal damage in all parts of the aorta regardless of the deployment catheter used.