Answers are grounded in this paper and cite their sources.
Reference tables covering the temperature range 0–1880°C have been established for platinum-40% rhodium/platinum-20% rhodium thermocouples. They are based on calibrations in air of ten thermocouples from four wire lots. These calibrations included freezing point determinations of zinc, antimony, silver, and gold; melting point determinations of palladium and platinum by the wire method; comparisons with standard platinum-10% rhodium/platinum and platinum-20% rhodium/platinum-5% rhodium thermocouples from 0 to 1750°C; and intercomparisons of the test thermocouples from 0 to 1850°C. The accuracy and stability to be expected from this thermocouple combination is discussed. More detailed tables than those presented here are available as APH Report No. 1292.
No takes yet. Share an insight, caveat, or question.
R. E. Bedford (1965) studied this question.