The amount of a sample compound extracted into a 1-μL drop of n -octane suspended in a stirred aqueous solution from the tip of a microsyringe needle is measured by gas chromatography (GC) as a function of time. The observed extraction rate curve is first order and yields the overall mass transfer coefficient for the sample compound, β o . For a given compound, β o varies linearly with stirring rate. Among the four compounds malathion, 4-methylacetophenone, 4-nitrotoluene, and progesterone, at a given stirring rate, β o is linearly proportional to the diffusion coefficient of the compound ( D aq ). This supports the film theory of convective−diffusive mass transfer, as opposed to the penetration theory. The relative standard deviation of the GC signal for 4-methylacetophenone after a 1.00 min extraction at 1500 rpm is 1.5%, which suggests that the system exhibits excellent potential as a tool for rapid analysis by solvent extraction/GC.
No takes yet. Share an insight, caveat, or question.
Jeannot et al. (1997) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: