Answers are grounded in this paper and cite their sources.
A study was made on the sensitivity of a magnetoresistive DNA-chip platform being developed for cystic fibrosis diagnostics. The chip, comprised of an array of 2.5×80μm2 U-shaped spin-valve sensors integrated within current line structures for magnetic label manipulation, enabled the detection at 30Hz of 250nm magnetic nanoparticles from 100pM down to the pM range (or a target DNA concentration of 500pM). It was observed that the sensor response increased linearly with label concentration. Noise spectra obtained for these sensors showed a thermal noise of 10−17V2∕Hz with a 1∕f knee at 50kHz at a 1mA sense current, showing that lower detection limits are possible.
No takes yet. Share an insight, caveat, or question.
Ferreira et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: