Key result
OMRON-MIT meets AAMI accuracy criteria with ~2 mmHg mean difference versus standard mercury sphygmomanometry.
Why the study?
The accuracy of the OMRON-MIT automated inflationary oscillometric blood pressure device needed evaluation against standard auscultatory mercury sphygmomanometry in the general population.
Does the OMRON-MIT automated inflationary oscillometric blood pressure monitor provide accurate measurements compared to standard auscultatory mercury sphygmomanometry in the general adult population?
Cross-Sectional
Blinded observers
Does the OMRON-MIT automated inflationary oscillometric blood pressure monitor provide accurate measurements compared to standard auscultatory mercury sphygmomanometry in the general adult population?
The OMRON-MIT inflationary oscillometric blood pressure monitor is accurate and meets standard validation criteria for clinical use in adults.
Answers are grounded in this paper and cite their sources.
Supports OMRON-MIT use in general populations; leaves open validation in special subgroups or settings.
BACKGROUND: To evaluate the accuracy of the OMRON-MIT, (an automated inflationary oscillometric blood pressure monitoring device) for use in the general population by direct comparison to standard auscultatory mercury sphygmomanometry. DESIGN: Two trained, blinded observers taking simultaneous blood pressure measurements alternating between a mercury sphygmomanometer and the OMRON-MIT. METHODS: The evaluation was carried out according to the revised British Hypertension Society protocol. The validation data was also classified by the American Association for the Advancement of Medical Instrumentation (AAMI). RESULTS: The OMRON-MIT device achieved a BHS grade A for both systolic blood pressure (SBP) and diastolic blood pressure (DBP). It also met the criteria for the AAMI protocol, the mean difference between the MIT and standard was -2.0 +/- 7 mmHg for SBP and -2.0 +/- 6 mmHg for DBP. CONCLUSIONS: Inflationary oscillometry is a valid technique for the measurement of blood pressure. The OMRON-MIT, which uses this technique, is accurate and can be recommended for clinical use in the adult population.
No takes yet. Share an insight, caveat, or question.
Golara et al. (2002) conducted a cross-sectional in Blood pressure measurement. OMRON-MIT automated inflationary oscillometric blood pressure monitoring device vs. Standard auscultatory mercury sphygmomanometry was evaluated on Accuracy of systolic and diastolic blood pressure measurement (MD -2.0 +/- 7 mmHg for SBP and -2.0 +/- 6 mmHg for DBP). The OMRON-MIT device met AAMI criteria with a mean difference of -2.0 ± 7 mmHg for SBP and -2.0 ± 6 mmHg for DBP compared to standard mercury sphygmomanometry.