Key result
Percutaneous ASD closure with the Occlutech device achieves 0% residual shunt at one year.
Why the study?
The long-term safety and efficacy of the Occlutech Figulla device for percutaneous atrial septal defect closure in adults had not been fully described.
Does percutaneous ASD closure using the Occlutech Figulla device provide safe and effective long-term outcomes in adults?
Cohort (n=166)
Does percutaneous ASD closure using the Occlutech Figulla device provide safe and effective long-term outcomes in adults?
Percutaneous ASD closure using the Occlutech Figulla device in adults is safe and highly effective at long-term follow-up, with no residual shunting at 1 year.
Answers are grounded in this paper and cite their sources.
Supports long-term safety and closure with Occlutech Figulla in adults; leaves open randomized trials for practice guidance.
PURPOSE: The Occlutech Figulla occluder has been proven safe and effective at midterm follow-up after percutaneous atrial septal defect (ASD) closure. We describe the safety and efficacy at long-term follow-up in adults. METHODS: All consecutive adult patients that underwent ASD closure between 2008 and 2015 were included. All complications were registered. Residual left-to-right shunt (LRS) was diagnosed using color-Doppler transthoracic echocardiography (TTE). Right-to-left shunting was diagnosed using contrast TTE. Successful closure was defined as no LRS at follow-up. RESULTS: In total, 166 patients (mean age 56.7 ± 16.1 years; 62% female) underwent percutaneous ASD closure using the Occlutech Flex I (70%) or Flex II (30%) device (diameter 24 mm; range 10-40 mm) under general anaesthesia and transoesophageal echocardiographic guidance. Long-term follow-up data were available for 144 patients (87%) with a mean follow-up of 5.9 ± 2.6 years, a total of 814 patient-years. During hospitalization, device embolization occurred in three patients (1.8%) with successful extraction in all. During the long-term follow-up, 15 patients (9.8%) suffered new-onset atrial fibrillation and stroke occurred in 2.1%. There was no residual LRS at 12-month follow-up. No device embolization occurred during the long-term follow-up. CONCLUSION: Percutaneous ASD closure using the Occlutech device appears to be safe at long-term follow-up with a high successful closure rate at one year.
No takes yet. Share an insight, caveat, or question.
Snijder et al. (2020) conducted a cohort in Atrial septal defect (ASD) (n=166). Percutaneous ASD closure using the Occlutech Figulla device was evaluated on Residual left-to-right shunt at 12-month follow-up. Percutaneous ASD closure using the Occlutech device in adults resulted in 0% residual left-to-right shunt at 12 months and a 1.8% device embolization rate during hospitalization.