We have developed an electrochemical biosensor for the detection of uranium, in the uranyl ion form, at low levels in water. This biosensor could have both academic and environmental monitoring applications. Uranium contamination of water sources poses serious health consequences, and this biosensor may aid in the improvement of targeting and the efficiency of remediation of contaminated water sources by improving the speed of initial identification as well as monitoring during remediation efforts. Our design uses a modified uranyl‐binding aptamer that has been inserted into an oligonucleotide scaffold to create a reusable, conformation‐switching electrochemical DNA biosensor with an apparent dissociation constant of 15±5 μM and a limit of detection within an order of magnitude of the Environmental Protection Agency's maximum contamination level for uranium in drinking water.