AIMS: The aim of this study was to develop a rapid assay for enumerating thermophilic bacteria in milk powder. METHODS AND RESULTS: The BactiFlow flow cytometer was used to count bacteria based on esterase activity in viable bacterial cells. A protocol for total viable bacteria was modified by heat-treating the sample to selectively label thermophilic bacteria. Samples of milk powder dissolved in 0.1% peptone were treated with 0.8% ethylenediaminetetraacetic acid to reduce background interference because of denatured milk proteins. Either thermophilic bacteria were added to the dissolved milk powder or milk powder solutions were incubated at 55 degrees C for 2-3 h to enrich the natural thermophile population for testing. Results from the BactiFlow were compared with traditional plate count results. CONCLUSIONS: Thermophilic bacteria in milk powder can be enumerated within 1 h using the BactiFlow flow cytometer. SIGNIFICANCE AND IMPACT OF THE STUDY: Microbiological test results obtained within 1 h can potentially be used to monitor manufacturing processes, effectively trace problems and provide confidence in the manufacture of product.
No takes yet. Share an insight, caveat, or question.
Flint et al. (2006) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: