This is the authors' abstract. We don't add key points for this paper.
In this article, we detail a comprehensive laboratory evaluation of an immunoassay for the rapid detection of ricin using the Meso Scale Diagnostics Sector PR2 Model 1800. For the assay evaluation, we used inclusivity, exclusivity, and informational panels comprised of extracts of 35 near-neighbor plant cultivar-extracts, 66 lectins, 26 white powders, 16 closely related toxins and proteins/toxoids, and a pool of 30 BioWatch filter extracts. The results show that the Meso Scale Diagnostics ricin detection assay exhibits good sensitivity and specificity with a limit of detection of 1.2 ng/mL. However, the dynamic range of the assay for the quantitation of ricin was limited. We observed a hook effect at higher ricin concentrations, which can lead to potential false negative results. A modification of the assay protocol that incorporates extra wash steps can decrease the hook effect and the potential for false negative results.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Manickam et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: