BackgroundOut-of-plane artifacts in digital breast tomosynthesis (DBT) can affect image quality, even subtly, and are influenced by the size and z-position of features with contrast of clinical images.ObjectiveTo propose a phantom and metric to further characterize out-of-plane artifacts in DBT.MethodsPhantoms with a signal inserted were manufactured, and the reconstructed planes were obtained using the DBT system. Normalized maximum contrast within the plane area was used to quantitatively evaluate out-of-plane artifacts. The spread of out-of-plane artifacts within the reconstructed plane was qualitatively evaluated by observing the profile within the plane area.ResultsThe larger the signal diameter, the stronger the effect of out-of-plane artifacts on the z-position far from the in-focus plane. When the z-position of the signal was on the upper side of the z-position of the center of X-ray tube rotation, out-of-plane artifacts were stronger on the upper side and weaker on the lower side of the signal. The spread of out-of-plane artifacts in the off-focus plane changed from monomodal to bimodal, with movement away from the signal's location in the z-direction.ConclusionsThis work proposes new phantoms and analysis methods to investigate the characteristics of out-of-plane artifacts, supplementing conventional methods.