Investigation of relative scintillation efficiency (so-called non-proportionality) and energy resolution of several recently developed solid-state organic scintillators is presented. The measurements were carried out at wide energy range between about 10 keVee and 4 MeVee using a Compton coincidence technique. High purity germanium detector was used as a reference to capture scatteredγ-rays at Compton electron energies below 1 MeVee. Above 1 MeVee, a large sizeLaBr₃: Ce crystal coupled to a photomultiplier was used as a reference detector to avoid neutron damage of germanium detector, as plutonium-beryllium neutron source was used to provide4438 keV γ-rays for Compton scattering. The signals were recorded in coincidence. By making narrow energy cuts on the reference detector energy axis, we evaluated the response of studied scintillation materials to nearly monoenergetic Compton electrons. Exemplary dataset presented for polystyrene shows continuous increase of non-proportionality from 0.67 at 12 keVee to 1.16 at 4 MeVee, when normalized to 1.000 at 477 keVee. Energy resolution in mentioned energy range changes from 73 % at 12 keVee down to 5.0 % at 4 MeVee. Statistical contribution dominates the energy resolution results at small energy depositions, while at high energies above 1 MeVee, intrinsic component plays crucial role. A set of basic scintillation parameters that influence materials performance will be presented for comparison.