Due to the lake of in-situ aerosol particle analysis systems, aerosol samples are taken and analyzed off-line. For detailed analysis of particle properties such as shape, morphology, and composition, off-line operating analytical tools like light microscopes, scanning electron microscopes (SEM), total reflection x-ray fluorescence (TXRF), and so on are used. The analysis must be performed on a representative sample of particles homogeneously deposited on a flat sample plate. This avoids sample preparation steps which may change the sample. In this paper we describe the design, construction, and evaluation of a continuous sampling device that deposits gasborne particles on an analytically suitable sample plate. The collection efficiency and the deposition pattern were optimized using a numerical model and experiments. It turned out that representative samples appropriate for further analysis can be taken in the particle size range from 0.03 mu m < Dp < 10 mu m. Additionally, the sampling efficiency was investigated for particles smaller than 0.03 mu m using electrical and non-electrical deposition mechanisms like diffusion and thermophoresis. The investigations performed demonstrate that the designed electrostatic precipitator (ESP) is a very useful tool for homogeneous particle deposition on analytically suitable flat sample plates and can be used as a back-up filter. Further, the ESP especially can be used in combination with a differential mobility analyzer (DMA) if detailed investigations of a narrow particle size range of a polydisperse aerosol are required.
No takes yet. Share an insight, caveat, or question.
Dixkens et al. (1999) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: